General Console Requirements; Console Environmental Requirements - GE LOGIQ P6 Series Service Manual

Table of Contents

Advertisement

GE H
EALTHCARE
5394152, Revision 5
D
IRECTION
Section 2-2

General Console Requirements

2-2-1

Console Environmental Requirements

Table 2-2
Temperature
Humidity
Pressure
Table 2-3
Power Source
Current Rating
Radiation Shielding
Heat Dissipation
Floor Landing
Floor Condition
2-2-1-1
Cooling
The cooling requirement for the LOGIQ™ P6/P6 Pro is 2000 BTU/hr. This figure does not include
cooling needed for lights, people, or other equipment in the room. Each person in the room places an
additional 300 BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and diameters can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interface.
Page 2-2
Environmental Requirements for LOGIQ™ P6/P6 Pro Scanners
Operational
o
10 - 30
C
o
50 - 86
F
30 - 85%
non-condensing
700 - 1060hPa
Environmental Requirements for an Ultrasound Room
Item
refer to Table 2-4 on page 2-3.
10.0A (100V-120V); 5.5A (220-240V)
NONE REQUIRED for ULTRASOUND ENERGY
Temperature
20-26 DEG. C (68-79 DEG F) for PATIENT COMFORT
Humidity
50% to 70% for PATIENT COMFORT
2000 BTU/Hr.
Approximately 680 - 800 kg/m
Gradient: WITHIN 5 degrees
Weight
80 kg (176lbs) without Accessories
Section 2-2 - General Console Requirements
LOGIQ™ P6/P6 P
Storage
o
-10 - 60
C
-40 - 60
o
14 - 140
F
-40 - 140
30 - 90%
30 - 90%
non-condensing
non-condensing
700 - 1060hPa
700 - 1060hPa
Values
2
without Accessories
S
RO
ERVICE
Transport
o
C
o
F
M
ANUAL