Section 2-2

General console requirements

2-2-1

Environmental Requirements

Table 2-1
Temperature
Humidity
Pressure
Heat Dissipation
2-2-1-1
Cooling
The cooling requirement for the LOGIQ S8 is 680 BTU/hr. This figure does not include cooling needed
for lights, people, or other equipment in the room.
Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2 - 2
System Environmental Requirements
Operational
o
o
10 - 35
C /50 - 95
F
30 - 80% (non-condensing)
700 - 1060 hPa
680BTU/hr
Section 2-2 - General console requirements
LOGIQ S8/LOGIQ E8 S
Storage
o
o
-10 - 50
C/14 - 122
F
10 - 90% (non-condensing)
700 - 1060 hPa
n/a
M
ERVICE
ANUAL
Transport
o
o
-10 - 50
C/14 - 122
F
10 - 90% (non-condensing)
700 - 1060 hPa
n/a