System Integration Checks

4-4-14-1
Operator panel Test
1.) Power ON the system.
2.) Check backlight on AN Keyboard, B Button, Freeze Key and TGC Knob.
3.) Select UTILITY in touch panel.
4.) Turn joystick knobs to adjust the backlighting of Touch Panel Light/Keyboard Light/Button Light.
5.) Verify backlight luminance (intensity) changes when knobs are turned.
PROPRIETARY TO GE
AN Keyboard
Figure 4-26 Operator panel
Utility
Touch Panel
Figure 4-27 Touch panel and knobs
Section 4-4 - Functional Checks
LOGIQ S8/LOGIQ E8 S
TGC Knob
B Button
Freeze Key
Button
Keyboard
M
ERVICE
ANUAL
4 - 35