Job Card Pm A155 - General System Inspection Checks - GE Senographe DS Manual

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1

Job Card PM A155 - General System Inspection Checks

1
SUPPLIES
None
2
TOOLS
None
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 10 minutes
4
SAFETY PRECAUTIONS
None
5
PREREQUISITES
None
6
PROCEDURE
6-1
Other General Inspection Checks
6-1-1
Visual inspection of cables
Check the appearance of all visible cables and correct hazardous trailing cables in the room.
6-1-2
Visual inspection of equipment
Note any signs of damage to covers, monitor support, etc.
6-1-3
Inspect the Radiation Screen
Check for breaks or cracks in the glass, gaps or cracks in the frame, and for correct, stable mounting,
etc.
6-2
System Cover Checks
6-2-1
Generator cover check
Check covers for security and presence of screw caps, and for damage, scratches, etc.
6-2-2
Gantry cover check
Check covers for security and presence of screw caps, and for damage, scratches, etc.
6-2-3
Control Station cover check
Check covers for security and presence of screw caps, and for damage, scratches, etc.