Brivo OEC 715/785/865 Mobile C-Arm X-ray Product Service Manual

8.6. System Function checks

Check the following areas of the C-Arm and workstation for proper operation.
8.6.1.

Perform Power On Checks

Observe safety procedures when operating motorized features.
WARNING
Perform the following steps with the system powered up and normally booted.
1. Connect the integrated cable in the C-Arm connector and lock it. Connect the Footswitch cable to the
C-Arm.
2. Plug the AC power cord into the receptacle.
3. Press the system on key located on the C-Arm control panel. Verify that the workstation and C-Arm
complete their software boot and initialization with no errors reported on the workstation and C-Arm
displays.
8.6.2.

C-Arm Checks

Mechanical Movement Test:
Refer to section 3.6.2 for C-arm mechanical movement checks.
If wigwag brake can't prevent Wig Wag motion effectively, or the lock and release
positions can't fit as the label indicating, perform below procedure to refresh the force of
wigwag brake.
38.5N<Force of wigwag brake <76.9N
Wigwag brake PM:
1. Pull up the handle far enough (about 3mm) that it clears the fixed shaft.
2. Turn the handle anticlockwise
3. Push down the handle.
8-10