General Console Requirements

2-2-1

Environmental Requirements of System based on Dell Precision 340

Table 2-2
Temperature
Humidity
Altitude
Table 2-3
Temperature
Humidity
Altitude
Table 2-4
Temperature
Humidity
Altitude
2-2-1-1

Cooling

The cooling requirement for the LOGIQworks is within the range 1220-2100 BTU/hr. This is depending
on whether a CRT Monitor or LCD Monitor is used, and on the hardware options installed. For detailed
BTU/hr. see table 2-7 and table 2-8. The BTU/hr. value relates to the maximum operational Power
consumption of the LOGIQworks Workstation. This figure does not include cooling needed for lights,
people, or other equipment in the room. Each person in the room places an additional 300 BTU/hr.
Demand on the cooling system.
2-2
5
EVISION
Dell Precision 340 Workstation
Operation
o
10 to 35
C
o
(50 to 95
F)
20% to 80%
non-condensing
-52 - 10000ft
-16 - 3048 m
NEC LCD 1880 Monitor
Operation
o
5 to 35
C
o
(41to 95
F)
30% to 80%
non-condensing
0 - 15912ft
0 - 4850m
Global Modem
Operation
o
0 to 50
C
o
(32 to 120
F)
20% to 90%
non-condensing
N/A
Section 2-2 - General Console Requirements
LOGIQ
Non-Operation
Storage
o
-40 to 65
C
-40 to 65
o
(- 40 to 149
F)
(- 40 to 149
20% to 80%
20% to 80%
non-condensing
non-condensing
-52 - 34777ft
52 - 34777ft
-16 - 10600 m
-16 - 10600 m
Non-Operation
Storage
o
-10 to 60
C
-10 to 60
o
(- 14 to 140
F)
(-14 to 140
10% to 85%
10% to 85%
non-condensing
non-condensing
0 - 45000ft
0 - 45000ft
0 - 13716m
0 - 13716m
Non-Operation
Storage
o
0 to 50
C
-20 to 70
o
(32 to 120
F)
(-4 to 158
20% to 90%
20% to 90%
non-condensing
non-condensing
N/A
B
S
WORKS
ASIC
ERVICE
Transport
o
o
C
-40 to 65
C
o
F)
(-40 to 149
20% to 80%
non-condensing
52 - 34777ft
-16 - 10600 m
Transport
o
o
C
-10 to 60
C
o
F)
(-14 to 140
10% to 85
non-condensing
0 - 45000ft
0 - 13716m
Transport
o
o
C
-20 to 70
o
o
F)
(-4 to 158
20% to 90%
non-condensing
N/A
N/A
M
ANUAL
o
F)
o
F)
C