Probe/Connectors Usage

4-4-9-1

Connecting a probe

1.) Place the probe's carrying case on a stable surface and open the case.
2.) Carefully remove the probe and unwrap the probe cable.
3.) Press the probe connector locking lever downwards.
4.) Press the connector locking lever upwards to the full vertical position to lock it.
5.) Carefully position the probe cord so that it is free to move and is not resting on the floor.
4-4-9-2

Activating the probe

Press the PROBE key on the alphanumeric keyboard to activate the "Probe Select" menu.
Select the appropriate probe using the TRACKBALL and the
Upon selection of an "Application", the programmed user presets appear.
Clicking a "Setting" field causes loading of the preset.
The probe is initialized, the main menu (2D mode) and the ultrasound image appears on the monitor in
write mode (real time display).
4-4-9-3

Deactivating the probe

When deactivating the probe, the probe is automatically placed in standby mode (read mode).
1.) Press the FREEZE key.
2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual of
3.) Carefully slide the probe around the right side of the keyboard, toward the probe holder.
4-4-9-4

Disconnecting the probe

Prior to disconnecting a probe freeze the image. It is unnecessary to switch the unit off.
If a probe is disconnected while running (write mode) a software error may occur.
CAUTION
In this case switch the unit OFF (perform a reset).
!! CAUTION:
1.) Press the locking lever downwards to unlock it.
2.) Carefully slide the probe and connector away from the probe port and around the right side of
4-4-9-5

Connecting a probe to the Dock Cart / Voluson Station

The Probe Box enables the connection of up to 3 probes to the Voluson i / Voluson e.
1.) Push the probe connector into the probe socket of the Voluson i / Voluson e.
2.) Lift the locking lever upwards to lock the probe connector.
10
EVISION
Afterwards align the connector with the probe port and carefully push it into place.
Voluson i / Voluson e for complete cleaning instructions.)
Ensure that the probe is placed gently in the probe holder.
the keyboard.
Chapter 4 - Functional Checks
V
OLUSON
S
.
TRACKBALL KEYS
i / V
e
OLUSON
M
ERVICE
ANUAL
4-25