Chapter 2. Installing the Pipe Nozzles
2.3.2 Selecting and Marking the First Nozzle Location - MR1 Mid-Radius (cont.)
3. Refer to Figure 29 below, and calculate the location for the center of the first nozzle as follows:
D 2W
------------------ -
×
Y
=
0
2
Note: See Table 3 on page 23 for tabulated AS/2 values calculated from the above equation for common pipe sizes.
Note: To clarify the use of the above equations, see the bottom of page 26 for a typical example calculation.