Determining And Marking The Second Nozzle Location - Mr2 Mid-Radius - GE T17 Installation Manual

Ultrasonic flow transducer
Hide thumbs Also See for T17:
Table of Contents

Advertisement

Chapter 2. Installing the Pipe Nozzles
2.4.2 Selecting and Marking the First Nozzle Location - MR2 Mid-Radius (cont.)
5. Spray the area around the first nozzle location mark with a marking dye product. Using a metal straight edge,
scribe 6" long vertical and a horizontal lines that intersect at the center punch mark (see Figure 52 below).

2.4.3 Determining and Marking the Second Nozzle Location - MR2 Mid-Radius

To determine and mark the second nozzle location, complete the following steps:
1. The position for the second nozzle is typically a distance equal to S along the pipe and located on the other side
of the pipe. For installation other than 45°, the distance is equal to S times the tangent of the installation angle.
Steps 2 to 7 on the following pages are for the second nozzle located at 45° from the first nozzle.
IMPORTANT: As shown above, while looking down onto the header pipe with the flow direction from left to right, the
second nozzle should be always located to the left of the first nozzle and the top reference point. The
nozzles cannot be physically mounted on the header pipe if they are mistakenly located at the mirror
image of the locations shown above.
42
Top of Pipe
Reference
Point
Figure 52: First Nozzle Scribe Lines
Second Nozzle Location
S
Top Reference Point
Figure 53: Calculating the Second Nozzle Location
S tan
(= S, when = 45 )
First Nozzle Location
Flow Direction
Model T17 Installation Guide

Advertisement

Table of Contents
loading

Table of Contents