Hitachi L44 Instruction Manual page 19

Probe
Table of Contents

Advertisement

Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
A) L44 probe
1) The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing detergent
is used, please also note the approved material compatibility for
the medical device.
2) Immerse the immersible part of the probe without connector into the
diluted detergent solution (see Fig. 5). Wipe the immersible part
of the probe under the surface of the detergent solution with a soft
cloth to remove all visible soil. Be sure that all grooves of the
probe are implemented during the cleaning process.
3) The immersible part of the probe should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
4) Wipe the un-immergible parts of the probe with a soft cloth dipped
with the detergent solution.
5) Rinse the probe with running tap water for 1 minute. (alternatively:
immerse the immersible part of the probe in a tray filled with
deionized water/tap water (see Fig. 5) for 5 min.)
6) Visually check the outer surface of the probe for cleanness. If
necessary, use magnifying glass for visually check. If there is still
soil visible, repeat all above steps.
B) Elasto coupler
1) The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing detergent
is used, please also note the approved material compatibility for
the medical device.
2) Immerse the elasto coupler into the diluted detergent solution. Wipe
it under the surface of the detergent solution with a soft cloth to
remove all visible soil. Be sure that all grooves of the elasto coupler
are implemented during the cleaning process.
3) The elasto coupler should be left in the detergent solution according
to the specified contact time of the detergent manufacturer.
4) Rinse the elasto coupler with running tap water for 1 minute.
(alternatively: immerse it in a tray filled with deionized water/tap
water (see Fig. 5) for 5 min.)
5) Visually check the outer surface of the elasto coupler for cleanness.
If necessary, use magnifying glass for visually check. If there is
still soil visible, repeat all above steps.
( 13 )
Q1E-EP1370