Siemens SITRANS RD500 Manual page 11

Configuring sitrans rd500 for connection to multiranger with modbus rtu
Hide thumbs Also See for SITRANS RD500:
Table of Contents

Advertisement

11. Click on Overview to return to the main page and confirm that values are updating.
12. Click on Logs. Download a .csv file under the Data Log section and confirm the correct
information is being logged.
There are many more features and options available in both the MultiRanger and SITRANS
RD500. Please consult the Operating Instructions for more information or contact your local
Siemens representative.
Configuring SITRANS RD500 for connection to MultiRanger with MODBUS RTU
Application examples, 09/2020, AG111710
3.3 Configuring RD500 to Read MODBUS Registers from MultiRanger
Planning/configuring
11