Phenomenon
ELECTROSTATIC
DISCHARGE
Radiated RF EM
fields
Proximity fields
from RF wireless
communications
equipment
RATED power
frequency
magnetic fields
a) The equipment is suitable for use in Home Health Environments and
Professional Health Care Environments limited to patient rooms and
respiratory treatment facilities in hospital or clinics. The more restrictive
IMMUNITY acceptance limits have been considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance of at least 15cm between
the ME EQUIPMENT or ME SYSTEM and sources of power frequency
magnetic fields.
20
Basic EMC standard
or test method
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
Important Information Regarding Electromagnetic Compatibility(emc)
Immunity test levels
Professional
HOME
healthcare
HEALTHCARE
facility
ENVIRONMENT
environment
± 8kV contact
± 2 kV, ±4kV ±, ±8 kV, ±15 kV air
10 V/m b) 80MHz
a)
- 2.7 GHz 80% AM
at 1kHz
COMPLIANT
NOTE: Further information about
distances to be maintained
between portable and mobile
RF communications equipment
(transmitters) and the Motorola
MBP66N can be requested
from AViTA using the contact
information provided in this manual.
However, it is advisable to keep
the electromechanical aerosol
equipment at an adequate distance
of, at least, 0.5 m from mobile
phones or other RF communications
transmitters to minimise possible
interference.
30 A/m c)
50 Hz or 60 Hz