4.8 Storage

Store the equipment in a cool, dustproof, dry, and dark space
to avoid high temperature, humidity and direct sunlight.
Limitations for the time for sterilized equipment belong to
package.

5. Maintenance and Safety Inspection

1) After using the probe, the probe should be cleaned and
disinfected and sterilized according to "4. Reprocessing
Procedure", store the probe in a cool and dark space to avoid
high temperature, humidity and direct sunlight.
2) Visually inspect the surface of the probe head, the housing, the cable
and the connector for any crack, scratch or denaturalization. If you
find any damage, do not use the probe, and immediately contact a
service support.