LG 14HK701G-W Owner's Manual page 15

Flat panel digital x-ray detector
Hide thumbs Also See for 14HK701G-W:
Table of Contents

Advertisement

Battery Charger
Item
Model
Size (Width x Height x Depth)
125.0 x 90.0 x 255.0 mm
Weight
Typ. 0.9 kg
Input
19 V
3.42 A
Output Nominal voltage
8.7 V
Battery Charger Adapter
Item
Model
Manufacturer
Asian Power Devices Inc. (APD)
Size (Width x Height x Depth)
134.0 x 59.8 x 31 mm
Weight
Typ. 0.3 kg
Input
AC 100-240 V~ 50-60 Hz, 1.5 A-0.7 A
Output
19 V
3.42 A
Classification by protection type
Class equipment
against Electric Shock
Cable length
1.5 m
Specification
Spec
Control Box
Item
Model
Size (Width x Height x Depth)
125.0 x 109.8 x 255.0 mm
Weight
Typ. 1.3 kg
Input
AC 100-240 V~ 50/60 Hz, 1.4-0.7 A
Output
24 V
2.1A, Trigger signals, Ethernet data for Detector A.
24 V
2.1A, Trigger signals, Ethernet data for Detector B.
Control Box supports 2 Detector connection.
Usage is, one is for Bucky stand, the other is for table (bed).
Generally, X-ray room of hospital installs 2 Detectors, Bucky stand and
table type, it's far more convenient and efficient working environment.
These 2 Detectors are not operated simultaneously, control box selects
the operating Detector by AWS command.
Transmission image/command between the Detector and PC.
Transmission control signals between the Detector and X-ray Generator.
Specification