Removing the lifting column

Removing the hand railing

Remove the left hand railing from the C-arm.

Removing the swivel brake

Remove the handle (1/Fig.9) and remove the plastic protector (2/Fig.9).
Remove the latch assembly by removing the screws.
Remove the cover plate under the swivel brake by removing the screws.
Knock the spring pin out and remove the bolt.
Move the C-arm into the horizontal position and secure it so that it cannot swivel.

Removing the C-arm

Lift the C-arm from its bearing support (three people required) and place it on a table
being careful not to damage the C-arm cable (Fig.10).
Remove the screws from the bearing shell of the C-arm bearing support.(1/Fig.11)
Move the lifting column all the way down by briefly applying 230 V
connection plug of the lifting column (N at X7 A/1, P at X7 A/3).
Removing the upper cover with the control panel
Remove the screws from the strain relief of the control panel cable (1/Fig.12).
Remove the protective conductor by removing the screws (2/Fig.12).
Remove the cover facing by removing the screws.
Remove the mounting screws (3/Fig.12).
Place the control lever at 90 degrees to the control panel.
Remove the cover ring around the control lever.
Loosen the cover (2/Fig.11) at the lifting column and remove it.
Remove the 4 mounting screws (Fig.13) on the lifting column.
Remove the upper cover with the control panel; to do this, lift the cover over the lifting
column and carefully pull it off over the control lever at the same time pulling the
connector of the control panel cable carefully through the rectangular opening (1/Fig.14).
Siemens AG
Medical Engineering
Register 3
SPR2-130.033.02
Rev. 02
Page 5 of 6
12.98
TD SD 24
2 - 5
at the 3-pole
AC
SIREMOBIL Compact
System Manual
2