Removing the lifting column
Fig. 1
Preliminary steps
•
Switch the SIREMOBIL off and disconnect the line voltage plug.
•
Secure the unit so that it cannot be inadvertently switched on.
•
To secure the unit against tipping, support it on the operating panel side with wooden
boards or other similar material.
Board D3
•
Remove the Plexiglas cover of board D3 by removing the screws.
•
Disconnect all plugs except D3/X2 (Fig.1).
•
Remove the protective conductor from the board.
NOTICE
Board D1
•
Disconnect all plugs from the board.
•
Remove the protective conductor from the board.
•
Remove all 6 mounting nuts (1/Fig.2).
•
Carefully remove the board and place it on an ESD surface.
•
Remove the spacers from the 6 screws.
•
Open the strain relief for ribbon cable X 4 (2/Fig.2).
Siemens AG
Medical Engineering
Board D3 does not have to be removed.
Register 3
SPR2-130.033.02
Rev. 02
2
Fig. 2
Page 1 of 6
12.98
TD SD 24
2 - 1
2
1
2
2
2
SIREMOBIL Compact
System Manual