X-ray flat panel imager

Advertisement

(2) Double Exposure
FPI Drive
Ready
EXP_RE
<IN>
EXP_OK <OUT>
EXPOSURE
DATA_STRAGE
FPI ENABLED
A: EXP_REQ width
Minimum 1ms
B: Refresh period
For line drive (2984 Line scan) 80 ms
C: EXP period
Set value: see the Image Acquisition mode table (500 to 2000 ms (by 500 ms step),
standard 500 ms).
D: Read period
For line drive (2984 Line Scan) 240 ms
①:EXP_REQ-EXP_OK period ........................................................................ Maximum 150 ms
②:Double EXPOSURE cycle period ................... Maximum 1700 ms (case EXPOSURE 500 ms)
B
C
D
Ref
Exposure
Read
1st
1st
A
C
1st image data
-7-
C
D
Ref
Exposure
Read
2nd
2nd
C
2nd image data
FDX3543RP