Download Print this page

Toshiba FDX3543RP Technical Data Manual page 9

X-ray flat panel imager

Advertisement

(2) Double Exposure
FPI Drive
Ready
EXP_RE
<IN>
EXP_OK <OUT>
EXPOSURE
DATA_STRAGE
FPI ENABLED
①:EXP_REQ-EXP_OK period ........................................................................ Maximum 150 ms
②:Double EXPOSURE cycle period ................... Maximum 1700 ms (case EXPOSURE 500 ms)
Ref
Exposure
Read
1st
1st
-9-
Ref
Exposure
Read
2nd
2nd
FDX3543RP