Rechargeable Battery; Rechargeable Battery - GE AED Pro Service Manual

Hide thumbs Also See for AED Pro:
Table of Contents

Advertisement

RECHARGEABLE BATTERY

The rechargeable battery (P/N 2023489-001) and charger (P/N 2023490-001) are separately sold accessories for the
Responder AED PRO.
DIRECTIONS FOR USE:
Model
2023681 Lithium
Sulfur Dioxide
If the yellow Charge LED blinks continuously, a charging error has occurred. Contact customer
service in the event of a charging error.
Rechargeable with LED
CAUTION: Use only Approved Equipment
The Rechargeable battery is made solely for Powerheart AED G3 Pro, and is NOT to be used with any other
AED models. Using batteries, pads, cables, or optional equipment other than those approved by the
manufacturer may cause the AED to function improperly during a rescue.
CAUTION: Lithium-ion Battery
Never short circuit, puncture, deform, or expose to temperatures above 65°C (149°F).
Page 16 of 33
Remove the rechargeable battery from the Responder AED PRO; the rechargeable
battery can only be recharged when removed from the Responder AED PRO.
Plug the charger into an appropriate electrical outlet.
Insert the charger cable into the rechargeable battery and ensure the yellow LED
above the rechargeable battery symbol is on. Charging is complete when the yellow
Charge LED goes out, and the four green Fuel Gauge LEDs are continuously lit.
Remove the charger cable from the battery when done charging. Charging may be
terminated early by removing the charger cable from the battery. If the battery is
charged for a minimum of 3 hours, the stated capacities will be met.
Estimated Shelf Life
5 Years
Insert Charger Here
Battery Capacity

Rechargeable Battery

(Also located on back)
© 2007 General Electric Company
Warranty
Typical Shocks
1 Year of 12 hours
of use, whichever
Up to 290 shocks
occurs first
Test Button