Electrical System Safety Checks; Features Selections - GE Dinamap PRO Series Manual

Hide thumbs Also See for Dinamap PRO Series:
Table of Contents

Advertisement

1.8 ELECTRICAL SYSTEM SAFETY CHECKS

1.9 FEATURES SELECTION
Revision A
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
The PRO Monitor is designed to protect the patient from electric
shock. To ensure the integrity of these safeguards, use a safety
analyzer to perform the following leakage current tests. These tests
should be performed on every unit whenever the case is opened
before it is returned to clinical service. If any reading exceeds the limit
specified, do not return the unit to clinical service.
Ground Leakage
Measure ground leakage current with normal polarity, reverse
polarity, and open neutral. Limit is 500 µA.
Temperature System Leakage (PRO 200 & 400)
Set your analyzer to measure leakage current with mains voltage
applied to the test probes. Use a temperature plug with the leads
shorted and apply the test probe of the analyzer to the leads. The
leakage reading should be less than 150 µA.
SpO
System Leakage (PRO 300 & 400)
2
Set your analyzer to measure leakage current with mains
voltage applied to the test probes. Use an SpO
leads shorted together and apply the test probe of the analyzer
to the leads. The leakage reading should be less than 150 µA.
The PRO Monitor allows the user to turn features (e.g., SpO
Printer) on and off. To access this option:
1. Power on the PRO Monitor and select the More... button.
2. Select Service mode and enter 3122.
3. Select the feature button you wish to change:
a. Select SpO
2
this button allows you to turn the SpO
the type of SpO
technology (Nellcor or Masimo) that is in your
2
Monitor to turn it on or off by using the rotor to select the option.
Once you have made your selection, a small check mark
appears next to the box that contains both a
Monitor does not have an SpO
be selected. Select OK and a tone sounds signalling that your
change has been made.
b. Feature Set Up: (Temp and Printer) Selecting this button
allows you to turn the Temp parameter and printer on or off by
using the rotor to select the option. Once you have made your
selection, a small check mark
2018554-001
: (NELLCOR, MASIMO, or NONE) Selecting
parameter on or off. Choose
2
parameter, the NONE box should
2
appears next to the box that
plug with the
2
, Temp,
2
and an" x." If your
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents