Chamber Temperature Setting; Key Lock Function - Panasonic MDF-137 Series Operating Instructions Manual

Biomedical freezer
Table of Contents

Advertisement

CHAMBER TEMPERATURE SETTING

Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
temperature is -25
Note: The unit is set at the factory that the set temperature -30
Table 1. Basic operation sequence (Example: Chamber temperature -25
Description of operation
1
Switch on the freezer.
2
Press set key.
Set the temperature to 25 with the
3
digit shift key and the numerical
value shift key.
4
Press set key.
The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
Although the value of the chamber temperature setting can range between -18
guaranteed temperature without load is -30

KEY LOCK FUNCTION

This unit is provided with the key lock function. When the key lock is ON, change of temperature setting
through the key pad is not available. The key lock is set in OFF at the factory.
Display
L 0
L 1
Table 2. Procedure for key lock setting (change from key lock OFF to key lock ON)
Description of operation
1
2
Press digit shift key for 5 seconds.
Press numerical value shift key and
3
scroll the figure to 1.
4
Press set key.
The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
o
C.
Key operated
Mode
Key lock is OFF
Key lock is ON
Key operated
o
C.
The current chamber temperature is
----
displayed.
The current setting is displayed and
SET
the second digit is flashed.
When pressed, the figure of settable
digit changes.
When pressed, the changeable digit
is shifted.
Set temperature is memorized and the
SET
current chamber temperature is
displayed.
o
C at ambient temperature of 35
Enable to change of temperature setting
Disable to change of temperature setting
The current chamber temperature is
----
displayed.
The current setting is displayed and
the first digit is flashed.
When pressed, the figure of settable
digit changes.
The key lock is set to ON.
SET
The current chamber temperature is
displayed.
14
o
C)
Indication after operation
o
C and -35
o
C.
Function
Indication after operation
o
C, the