Functional and electrical safety checks: Functional Checkout procedures
4. From the BIS parameter menu, select the SENSOR CHECK menu
option. An information window opens with the impedance data. For
example:
5. Verify that the impedance values displayed in the information
window fall within the range shown below.
Dash 3000/4000/5000
LEAD
IMPEDANCE
1
4.8
2
14.0
3
2.3
4
3.3
Acceptance range
Electrode #
in Kohms
1
4—6
2
8—17
3
2—4
4
3—5
STATUS
PASS
PASS
PASS
PASS