MN1-6119 Rev. 3

2-2. Name of each parts

The name of each part is shown in Figure 2 and the explanation for each part is listed in Table 1.
Ultrasonic radiation part
Front side
Probe tip
Cable
Connector
Caution
Do not pull, bend, twist, or apply excessive force to the cable.
The probe may malfunction due to cable disconnection.
Do not subject the ultrasonic radiation part to hard impact.
The impact may cause damage to the transducer, and that results in noise or no echo in the image.
In most cases, the ultrasonic radiation part itself is not damaged because the part is made of elastic
material.
Ultrasonic radiation part
Front side
Probe tip
Figure 2 Name of each parts
Table 1 Name of each part and its explanation
Name
Connector
Explanation
Ultrasound is radiated from this part. The electronic
convex transducer is integrated underneath this part.
The side corresponds to the side of the orientation mark
on the image.
This part is gripped during operation.
Cable transfers electric input/output signals.
The connector is the part which is connected to the
ultrasound diagnostic instrument.