Reynolds Correction - GE DigitalFlow DF868 Programming Manual

Panametrics multipurpose ultrasonic liquid flowmeters
Hide thumbs Also See for DigitalFlow DF868:
Table of Contents

Advertisement

NO
Additional Selections
WATER
Enter the Water Temperature and
press [ENT].
MIXED
Enter the Fluid Soundspeed and
press [ENT]. Then enter Percentage
of Water and press [ENT].
OTHER
Enter the Fluid Soundspeed and
press [ENT].

Reynolds Correction

Programming Site Data
Table 1-10: Fluid Types for ENERGY ON
Tracking Windows =
W260
MIXED
OTHER
[F1]
11. Press
to turn Reynolds Correction off, or
Note: Reynolds Correction is a number based on the Kinematic
Viscosity and flow rate of the fluid. It should be enabled for
most applications. Also, if you have enabled a temperature
compensation table (see page 1-55), a message indicates that
the table is active.
12.When you enable the Reynolds Correction Factor, you must also
enter the Kinematic Viscosity of your fluid, as listed in Sound
Speeds and Pipe Size Data. Use the numeric keys to enter a value,
[ENT]
and press
.
YES
Additional Selections
No additional selections required.
Enter the Percentage of Water and
press [ENT].
Enter the Minimum Soundspeed and
press [ENT]. Then enter Maximum
Soundspeed and press [ENT].
March 2008
[F2]
to turn it on.
1-19

Advertisement

Table of Contents
loading

Table of Contents