Drying; Inspection - Hitachi EUP-F531 Instruction Manual

Finger-top probe
Table of Contents

Advertisement

Manual disinfection:
1) Prepare the disinfectant solution in a tank with cold water (please
follow the instructions of the disinfectant manufacturer regarding
application, concentration, microbiological efficiency, service
life and contact time).
2) Confirm the concentration of the disinfectant before immersing the
probe. Although Cidex® OPA does not need to be diluted, it is
recommended to use test strips to verify the concentration. The test
strips can indicate whether or not the concentration is above the
Minimum Effective Concentration (MEC). Please also note the
expiration date of the test stripes. Temperature of disinfectant
solution should be minimum 20 °C. The minimum contact time is 5 minutes.
If a different disinfectant is used, follow the manufacturer's
instructions. Please also consider the material compatibility for
the medical device.
3) Immerse the immersible part of the probe into
the disinfectant (see Fig. 3). Set a clock to
insure the recommended contact time which is
5 minutes.
4) Rinse the immersible part of the probe with
deionized water for 1 minute. (alternatively:
immerse the immersible part of the probe in
a tray filled with deionized water (see Fig.
3) for 5 min.)
5) Visually check the outer surface of the probe
for
leavings
necessary, repeat the rinsing.
4.4.

Drying

1) Wipe the equipment with a single-use, fluff-free wipe
or towel to remove moisture from the surface of the equipment.
2) Dry the equipment naturally in an ambient temperature between 15-30°C
for a minimum of 4 hours. Alternatively the equipment can be dried
using a drying heater at a temperature of less than 55 °C.
4.5.

Inspection

Inspect the equipment for any damage such as crack, scratch or
deformation. Do not use it if any damage is found.
of
the
disinfectant.
-8-
Water, Detergent
or Disinfectant
If
Fig. 3
Immersion of
the probe
Q1E-EP0415
Drying

Advertisement

Table of Contents
loading

Table of Contents