Activating A Channel/Path - Chx-Activ; Accessing The Activ Submenu - GE AquaTrans AT868 User Manual

Ultrasonic flow transmitter for water (1- & 2-channel)
Hide thumbs Also See for AquaTrans AT868:
Table of Contents

Advertisement

September 2004
Activating a Channel/
Path - CHx-ACTIV
Accessing the ACTIV
Submenu
Procedure Options
2-2
ACTIV
The
submenu is used to activate/deactivate the channel/path.
The channel/path should be activated when you receive your unit;
however, you should verify that the channel/path is active before you
begin programming.
While following the programming instructions, refer to Figure A-1 on
page A-1 in Appendix A, Menu Maps. Remember to record all
programmed data in Appendix C, Data Records.
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [ ] until PROG appears and press [ENTER].
3. Press [ ] until the desired channel appears. Press [ENTER] at CH1
or CH2.
4. Press [ ] until ACTIV appears and press [ENTER].
5. Press [ ] until TRANS appears and press [ENTER]. Selecting
TRANS activates the channel/path. If you have a 2-channel/path
unit, you can also disable a channel/path. If you select OFF you
will not be able to program that channel/path.
After completing the above step, the user program returns to the
Channel PROGRAM window. Do one of the following:
To continue entering "quick startup" data, press [ESC] once and
proceed to Step 3 in the following section.
To continue regular programming, refer to Appendix A, Menu
Maps, to navigate to the desired menu.
To leave the user program, press [ESC] three times.
Programming Site Data

Hide quick links: