incompatible with the system and could damage the scanner.
Follow the manufacturer's instructions, recommendations, and guidelines for
cleaners and disinfectants, as well as your regional regulations.
Check expiry dates, concentration, and efficacy of the chemicals used.
Wear the appropriate personal protective equipment (PPE), such as eyewear and
gloves, as recommended by the chemical manufacture.
Repeated use and cleaning over the course of the scanner's life may deteriorate
its cleanliness.
Using incompatible solutions to clean the scanner may damage its surface.
The scanner and its parts (including accessories) may not withstand the cleaning
or disinfecting processes (including repetitive process) specified in this manual
and may damage or deteriorate its safety provisions .
Cleaning or disinfecting the scanner while the battery is charging may cause the
battery to short-circuit and overheat, causing an electric shock or burn.
Cleaning or disinfecting the scanner using other than IPA (isopropyl alcohol) may
damage it.
During an emergency where the scanner is used to examine multiple patients in a short
period of time, the lack of proper cleaning and disinfecting between patients may
spread infections to other patients and users.
Recommendations for cleaning the ultrasound probe as following step:
Turn off your devices before cleaning it.
To be ensured that all the coupling gel and other visible substances from the
probe is removed by wiping with a clean paper towel. If necessary, to remove
material dried to the surface, the cloth can be moistened with lukewarm water.
It shall inspect the probe's lens and casing after each use. To check out any
damage that would allow liquid to enter the probe. If the user found a probe
damage, the probe shall not be placed into any liquid (e.g., for disinfection) and
shall not be used until it has been inspected and repaired/replaced by ASUSor a
local distributor for service.
Recommendations for disinfecting the ultrasound probe (After cleaning):
Spray 70% Isopropyl Alcohol onto the surface of probe head.
Repeat step one for two or three times.
Wipe out the disinfectant with a clean paper towel.

Maintenance