Scheduled Maintenance; Recommended Spare Parts; Troubleshooting; Figure 2-6: Dbdid Drawing - ABB PGC5000 Service Instruction

Gas chromatograph
Hide thumbs Also See for PGC5000:
Table of Contents

Advertisement

NOTICE – Equipment damage. Take great care to avoid any leaks associated with this
detector, as leaks can cause a variety of problems. See Troubleshooting in section 3.
2.4

Scheduled maintenance

There are no parts within the detector power supply or electrometer which require scheduled
maintenance.
2.5

Recommended spare parts

Part Number
Description
852A038-1
Power Supply DBDID
885A005-1
Detector Cell Assembly DBDID
851K006-1
FID/DBDID Amplifier Kit
(Order 851K006-1 Amp Kit to build the standard
851A094-11412 per included 851J003 bulletin.)
Contact your local ABB sales and service representative for specific instructions for ordering spare
parts. See the last page of this instruction for contact information. Always include the information
listed in Equipment Identifications and Configuration Identification in your request.
When ordering parts for replacement, use the list of parts included in the Engineering Data Package
that was provided with the analyzer to insure the correct version of each part.

Figure 2-6: DBDID drawing

3

Troubleshooting

This section describes general troubleshooting procedures, issues, and remedies.
Before proceeding with further troubleshooting procedures, first verify:
 The system has power.
 All switches and gas settings are correct.
 Column flow rates and pressure settings are correct.
 Using a VOM meter, verify the voltage between chassis ground and the test points on the
bottom of the power supply board are as follows:
+10 volts DC Bias voltage
+2.5 volts DC Feedback voltage
+2.5 volts DC Plasma voltage
Quantity
1
1
1
89 2J 00 9MN A E | D B DI D | 9