XCU Download POLYDOROS SX 65/80 ......9 - 1 10 ______Exiting the System Configuration ________________________________ 10 - 1 11 ______Changes to Previous Version____________________________________ 11 - 1 Siemens AG RXL2-120.034.05 Page 3 of 4 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 4
0 - 4 Contents This page intentionally left blank. LITHOSTAR Multiline RXL2-120.034.05 Page 4 of 4 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Prerequisites and Remarks For the tools and test equipment required, refer to the respective NOTICE LITHOSTAR Multiline Start-up Instructions... • Prerequisites for configuring the system using the Service PC are: - Basic knowledge such as PC handling and operation of the Windows program.
Select components in the "available" field located in the "Configure:Site Structure:Component selection" window. Window name Field Window Window Message window • Exit the window with OK. LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 4 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Page 7
Important: - Connect the system controller (D320.XCS) to Com 1. - Connect the PC mouse to the PC BUS mouse connector. Siemens AG RXL2-120.034.05 Page 3 of 4 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
AN 0 OUT A AN 1 OUT B AN 2 I_ANL 125V_FL AN 3 9 10 AN 4 KURZ_AN AN 5 LITHOSTAR Multiline RXL2-120.034.05 Page 4 of 4 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Page 9
Download (load software onto generator) Chap.8 Exit system configur- ation (System/Quit) Chap.9 Siemens AG RXL2-120.034.05 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 10
Help Text • Select "Help/Index" in the main menu. A help text can be called up for every active window by pressing NOTICE the F1 key. LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 2 Siemens AG System Manual Rev. 02 12.99...
If "Demo" is checked off, the software can work without the XCU. In the "Demo function", communication with the XCU is not possi- ble. • Exit the window with OK. Siemens AG RXL2-120.034.05 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 12
A help text can be obtained for every active window by pressing the F1 key. If the password field is empty, obtain the password from the last Speed info with the title "Password Speed info". LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 2 Siemens AG System Manual Rev.
Help texts for configuration are available by pressing the F1 key. If a system-tested system has crashed, download the configuration from the back-up floppy. Use the "Data Restore" menu. Siemens AG RXL2-120.034.05 Page 1 of 12 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 14
4 - 2 System Configuration LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 12 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Page 15
SID Tracking not available SDM Flip not available max tubecurrent < 100 ms 1000 mA at VD02A 800 mA at VE00J Siemens AG RXL2-120.034.05 Page 3 of 12 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 16
4 - 4 System Configuration The configuration will be interupted unless all necessary components are selected. LITHOSTAR Multiline RXL2-120.034.05 Page 4 of 12 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Page 17
System Configuration 4 - 5 Location of the symbols varies depending on software version. Siemens AG RXL2-120.034.05 Page 5 of 12 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 18
150/40/80/HC-10xL" Select "Multiline Memoskop" Click on "F" Select "SDM (spotfilm)" Depending on the SW version, additional symbols are available in the form. LITHOSTAR Multiline RXL2-120.034.05 Page 6 of 12 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Page 19
If you wish to set different dose values in zoom mode, you must adjust the zoom ratios accordingly. Depending on the software version, different symbols are available. Siemens AG RXL2-120.034.05 Page 7 of 12 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 20
The "Mode Selection" field varies depending on the software. With VE00J, "Normal Fluoro" is designated "Cont. Fluoro". Select "Litho 63 KV" Enter "113" (not with VE00J) LITHOSTAR Multiline RXL2-120.034.05 Page 8 of 12 Siemens AG System Manual Rev. 02 12.99 TD PS 24...
Page 21
System Configuration 4 - 9 Select Litho 63 kV HK Enter "113" (not with VE00J) Siemens AG RXL2-120.034.05 Page 9 of 12 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 22
PC. Enter "Region 1" Enter "Organ 1" Text varies depending on the software version. Depending on the software version, additional symbols are visible. LITHOSTAR Multiline RXL2-120.034.05 Page 10 of 12 Siemens AG System Manual Rev. 02 12.99...
Page 23
Notice: For more see 4-10 Depending on the software version, the text and symbols are arranged differently in this field. Siemens AG RXL2-120.034.05 Page 11 of 12 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 24
System Configuration 6WDUWXS Working on: ...... Enter freely selectable text here, e.g. “Start-up”. This can be read out in the “Main Program / Info”. Generator OFF / ON LITHOSTAR Multiline RXL2-120.034.05 Page 12 of 12 Siemens AG System Manual Rev. 02 12.99...
"OK Button" is active. First the small focus and then the large focus is calibrated (condi- NOTICE tioned). • Exit the window with OK. Siemens AG RXL2-120.034.05 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Exit the window with OK. NOTICE The data may be updated in the XCU with "Transfer Value" allowing test exposures to be made without exiting the service software window. LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 2 Siemens AG System Manual Rev.
Calculate the dose at the I.I. input (K ) as follows: Q g2 Q g2 K Tg × × ---------- - ---------- - ---------- - Q g 1 Siemens AG RXL2-120.034.05 Page 1 of 4 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 28
I.I. plane 0,94 0,94 × × Measurement chamber be- K T g ------------------ - ------------------ - 0,94 0,94 fore attenuating elements LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 4 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Refer to Chapter 6 of the Start-Up Instructions, RXL2-120.034.., for settings/tests. For POLYDOROS SX with optional instant stored image Setting the dose refer to chapter 6, Start-up instructions RXL2-120.034..Siemens AG RXL2-120.034.05 Page 3 of 4 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Adjusting the Maximum Skin Dose Rate NOTICE Perform this adjustment only in those countries which mandate a limitation of the skin dose rate. Refer to Chapter 6 of the Start-Up Instructions, RXL2-120.034.., for settings/tests. LITHOSTAR Multiline RXL2-120.034.05 Page 4 of 4 Siemens AG System Manual Rev.
Page 31
Select Adjustment Results in the "Polydoros .. Service/Diagnostics/" window. CAUTION Press the F1 key and read the help text before performing any steps. Siemens AG RXL2-120.034.05 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 32
“max. 8 characters” err e.g. elog01.err or elog02.err You can save the file to the hard drive or to the diskette. Press the "More Info" key for further information about troubleshooting. LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 2 Siemens AG System Manual Rev.
Page 33
Insert the back-up diskette in Drive A: . • Select Backup to disk in the "XCS Service Application - Main Program/Data/" window. Siemens AG RXL2-120.034.05 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 34
8 - 2 Saving the Data on Diskette (Backup) This page intentionally left blank. LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 2 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Page 35
- the red LED on D220 is flashing - the software has crashed A download is not necessary after the D220 is replaced. Siemens AG RXL2-120.034.05 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 36
Select Download to Unit in the "XCS-Service Application - Main Program/Date/" window. • Select SX Heating in the "Data/Download select Component" window. • Select the Start Button in the "Data/Download" window. LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 2 Siemens AG System Manual Rev. 02 12.99...
Page 37
(Texts can be read out in “Main Program / Info”). • Switch the generator OFF / ON. CAUTION Perform a “Backup” after the adjustment has been completed. Siemens AG RXL2-120.034.05 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 38
10 - 2 Exiting the System Configuration This page intentionally left blank. LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 2 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...
Page 39
Page 4-11 * Text newly included Chap. 5 - 2 Para: DIAMENTOR MS alignment..deleted TD PS 24/ Erlwein TD SP/ M. Heinlein SMS / Iselin Siemens AG RXL2-120.034.05 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 12.99 TD PS 24 System Manual...
Page 40
11 - 2 Changes to Previous Version This page intentionally left blank. LITHOSTAR Multiline RXL2-120.034.05 Page 2 of 2 Siemens AG System Manual Rev. 02 12.99 TD PS 24 Medical Engineering...