GE B20 User Manual page 146

Patient monitor
Hide thumbs Also See for B20:
Table of Contents

Advertisement

Nellcor SpO
specifications
2
Measurement range
Display range
Calibrated against functional oxygen saturation.
Measurement accuracy
Display resolution
Display averaging
Pulse rate
Measurement and display
range
Display resolution
Measurement accuracy
Sensor Light Source*
Wavelength
Total optical output power
of the sensor LEDs
*
This information may be useful to clinicians, such as those performing
photodynamic therapy.
1 to 100%
0 to 100%
Adult
100 to 70% ±2 digits
Neo
100 to 70% ±3 digits
Low perfusion 100 to 70% ±2 digits
1% of SpO
2
2 to 7 seconds
20 to 250 bpm
1 bpm
±3 digits
Infrared: 900 nm
Red: 660 nm
less than 15 mW
142
Masimo SpO
specifications
2
Measurement range
Display range
Calibrated against functional oxygen saturation.
Measurement accuracy
Without motion
With motion
Low perfusion
Display resolution
Display averaging
Pulse rate
Measurement and display
range
Display resolution
Measurement accuracy
Sensor Light Source*
Wavelength
Power Dissipation
1 to 100%
0 to 100%
Adult/Pediatric 100 to 70% ±2 digits
Neonate
100 to 70% ±3 digits
Adult/Ped/Neo 100 to 70% ±3 digits
100 to 70% ±2 digits
0~69% unspecified
1% of SpO
2
2 to 16 seconds
25 to 240 bpm
1 bpm
Without motion
±3 bpm
With motion
±5 bpm
Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Infrared: 22.5 mW (max)
Red: 27.5 mW (max)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

B40

Table of Contents