Circumference/Area Measurement; Two Distance Method - GE LOGIQ A100 Technical Publication

Table of Contents

Advertisement

Circumference/Area Measurement

Circumference and area can be measured by three
methods:
D
D
D

Two Distance Method

Scan to obtain a section of the anatomical organ whose
circumference/area is to be measured.
Press FREEZE to stop image acquisition.
A "
Press the MEASUREMENT key once to display the
Meas
General Measurement Menu. An open ended " " cursor
appears on the screen.
Press 1
T he c ircumference and area will be calculated and
1
displayed after completion of two measurements.
Use the TRACKBALL to move the open ended " " cursor
to either end of the first axis of the area to measure.
TB
Press SET to fix the start point cursor. Distance in mm
appears on the left side of the screen.
continued
t
α 100 User Manual
LOGIQ
2211157–100 Rev 0
two distance method
ellipse method
trace method.
Basic Measurements
105