Acquiring 4D Aortic valve using Enface Left/ Right
The Enface Left/Right Flexi Zoom buttons allow you to quickly view the image displayed in either the
left or right 4D zoom plane in 4D – to access press either Enface Left or Right once you have optimized
your ROI boxes.
Note if you acquire a long axis you can change to a short axis live or post acquisition by using the Angle button.
Adding more definition to that AV 4D dataset
The first thing to consider with your aortic valve dataset is your gains.
A darker 2D Gain will produce a clearer 4D dataset.
When in 4D your Active Gain knob becomes 4D gain.
To add more 4D definition to your structure turn the Active Gain knob down a notch or two.
It may also be necessary to adjust the Translate line to optimally display 4D structures.
JB83797AU
13