Installation and setting
Fig. 1

Unpacking, applying warning labels

?
WARNING
Fill in the LINA cards and return them to the addresses stated in the instructions on the
cards.
Unpack the protractor.
Connect the exposure release cable.
Mount warning label 60 77 411. Choose appropriate language version and paste it/them
on the control panel according to X, Y/fig. 1.
Only in Germany:
Mount label 96 33 645 on the lower arm 50 mm above the link, see fig. 2.
NOTICE
Check the unit for cracks, fractures or other damages.
Make sure that neither the single tank nor the collimator has any mechanical defects
which would impair the radiation protection.
Siemens-Elema AB
Solna, Sweden
In order to avoid serious damages due to humidity condensation
or to high temperature inside the batteries, assure that the unit
has established normal room temperature. The single tank tem-
perature must be >15°C before startup.
If, due to increased operating loads (DIN 6815), there are other
dimensions on site for the control range, appropriate labels can
be ordered:
- Control range: 2.5 m, Part No. 84 14 823
- Control range: 3.5 m, Part No. 31 36 731
SPR8-220.033.01
Rev. 05
Fig. 2
Page 1 of 10
05.01
SPS-UD
2 - 1
2
50 mm