Check The Arm- And Single Tank Movement; Adjustment Upper Kv / Mas Limits - Siemens MOBILETT Plus HP Installation And Setting Instructions Manual

Table of Contents

Advertisement

Installation and setting
Set the following exposure parameters and release exposure:
kV Meter
selection
DC voltage
* The measurement inaccuracy of the respective measurement device must be sub-
tracted from these limit values.
Connect the power plug and switch the main switch to position "M"
Set the following exposure parameters and release exposure:
kV Meter
selection
DC voltage
Disconnect the power plug, roll up the cable and switch the main switch to position "B".

Check the arm- and single tank movement

?
Check the arm systems and the single tanks up and down movements. The arm and the
single tank should be easy to move and remain in desired positions. If this is not the
case, see "Service Instructions, section Adjustment of the friction linings".

Adjustment upper kV / mAs limits

If the customer or local regulations asks for a limitation of the upper kV / mAs value, the
service program P14 and / or P15 have to be executed.
Select Service Mode according to chapter "Capacitor formation".
Select P14 (for upper kV limit) or
Select P15 (for upper mAs limit).
Press lamp button to display actual kV limit / mAs limit.
Press kV / mAs +/- button to change value.
Press lamp button to store value and leave program.
Switch "1" of S1 on D1 to "OFF".
Check the chosen values.
Siemens-Elema AB
Solna, Sweden
Exposure parameter
52 kV
50 mAs
81 kV
20 mAs
133 kV
12.5 mAs
Exposure parameter
81 kV
20 mAs
SPR8-220.033.01
Rev. 05
Limit values* to be maintained
49.0 - 55.0 kV
77.0 - 85.0 kV
126.4 - 139.6 kV
Limit values* to be maintained
77.0 - 85.0 kV
Page 7 of 10