Ip Removal Unit (Cleaning The Rubber Rollers - Siemens FUJI-Reader Service Manual

Table of Contents

Advertisement

7.6

IP Removal Unit (Cleaning the Rubber Rollers)

1Y
(1) Clean the rubber rollers.
(2) By reversing the procedures described in "7.5 IP Removal Unit (Cleaning/Replacing
the Shock Absorber-Attached Guide Plates in the Conveyance Path of the IP Removal
Unit)," reinstall the shock absorber-attached guide plates A, B, C, and D.
Anti-static member
Moistened cloth
CHECK
Make sure that the anti-static members are brought into contact with the rollers.
009-058-02
009-058-00
08.30.2001 FM3142
10.20.2000 FM2732 (2)
At intervals of one year, or when 60,000 IPs have been processed
CHECK
Rubber roller
[Clean]
1Y
CR-IR347 Service Manual
CHECK
Anti-static member
IP removal unit
PM - 38
Clean.
Anti-static member
CHECK
FR7H6151.EPS
PM - 38