Restore System Setups
Restore Options
Revision E
Maintenance: Disassembly Guidelines
the label of the system.
6.
Select/verify that No is selected for Update Primary Boot.
7.
Enter the Print head resistance. This number can be found on the print head
label.
8.
Select the appropriate language in the Keyboard menu.
9.
Select Return.
1.
Power up the cart.
2.
From the Main Menu, select System Setup.
3.
Enter the System Password and press Enter.
4.
Select Restore Setup from the System Setup menu.
5.
Select From SD Card from the Restore Setup menu.
Using the option activation codes for the system, restore the options which had been
installed on the board which was removed. These options are printed on a label
located on the bottom of the paper tray.
NOTE
Use the activator codes shown on the label on your system. The activator codes
shown in the figure below are examples only.
1.
Power on the cart.
2.
Within the System Setup function, select Basic System.
3.
Select Option Activation to activate options.
4.
Type the 12-digit option activation code and press the Enter key.
5.
Repeat the previous step for each option to be activated on the new system.
MAC 5500 resting ECG analysis system
2020299-020
122A
5-17