Siemens Healthineers ADVIA Quick Reference Manual page 14

Autoslide system

Advertisement

Weekly Shutdown - Startup
1. Select Procedures > Clean Autoslide >
Weekly Shutdown
2. Follow the on screen instructions. Remove
stain and buffer lines from reagent bottles
and place in empty containers. Use separate
containers for buffer and stain lines.
3. Select Next; the system will perform a 5
minute air prime.
4. When prompted, move buffer and stain lines
to containers containing methanol. Use a
separate container for buffer and stain lines.
5. When complete, Exit Weekly Shutdown and
select Weekly Startup.
6. Follow the on screen instructions. Place stain
and buffer lines in the empty containers and
select Next.
7. When prompted, place lines in the reagent
bottles. Select Next.
8. Select Exit when Weekly Startup complete.
14