Autoslide system

Advertisement

Printer Ribbon Replacement
1. Analyzer Mode: Standby
Autoslide Mode: Off
2. Open the Autoslide lower front cover and remove
the slide rack loader.
3. Remove the printer ribbon by holding both sides
of the printer cartridge and press towards the
inside of the Autoslide. Pull the ribbon cartridge of
the holder.
4. Install a new ribbon in the holder. Verify the
notches on the ribbon cartridge are positioned
correctly in the holder.
5. Replace the slide rack loader. Close the lower front
door of the Autoslide.
6. Return the analyzer to Ready to Run and turn on
the Autoslide.
7. The Autoslide will automatically perform a
Mechanical Initialization. Allow the initialization to
complete before using system.
17