GE B40 Reference Manual page 27

Patient monitor
Hide thumbs Also See for B40:
Table of Contents

Advertisement

Sample Messages: Transactions from the monitor HL7 interface
PV1||E|X^^B4020
OBR|1|080019FFFE0B402020121109160900^VSP^080019FFFE0B
4020^EUI-
64|080019FFFE0B402020121109160900^VSP^080019FFFE0B402
0^EUI-64|182777000^monitoring of
patient^SCT|||20121109160900
OBX|1||69965^MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS^MDC|1
.0.0.0|||||||X
OBX|2||69854^MDC_DEV_METER_PRESS_BLD_VMD^MDC|1.13.0.0
|||||||X
OBX|3||69855^MDC_DEV_METER_PRESS_BLD_CHAN^MDC|1.13.1.
0|||||||X
OBX|4|NM|150033^MDC_PRESS_BLD_ART_SYS^MDC|1.13.1.1|11
2|266016^MDC_DIM_MMHG^MDC|||||R|||||||080019FFFE0B402
0^B40_GE
OBX|5|NM|150034^MDC_PRESS_BLD_ART_DIA^MDC|1.13.1.2|76
|266016^MDC_DIM_MMHG^MDC|||||R|||||||080019FFFE0B4020
^B40_GE
OBX|6|NM|150035^MDC_PRESS_BLD_ART_MEAN^MDC|1.13.1.3|9
5|266016^MDC_DIM_MMHG^MDC|||||R|||||||080019FFFE0B402
0^B40_GE
OBX|7|NM|149522^MDC_BLD_PULS_RATE_INV^MDC|1.13.1.4|80
|264864^MDC_DIM_BEAT_PER_MIN^MDC|||||R|||||||080019FF
FE0B4020^B40_GE
OBX|8||69855^MDC_DEV_METER_PRESS_BLD_CHAN^MDC|1.13.2.
0|||||||X
OBX|9|NM|150087^MDC_PRESS_BLD_VEN_CENT_MEAN^MDC|1.13.
2.1|9|266016^MDC_DIM_MMHG^MDC|||||R|||||||080019FFFE0
B4020^B40_GE
OBX|10||69798^MDC_DEV_ECG_VMD^MDC|1.5.0.0|||||||X
OBX|11|NM|147842^MDC_ECG_HEART_RATE^MDC|1.5.1.1|80|26
4864^MDC_DIM_BEAT_PER_MIN^MDC|||||R|||||||080019FFFE0
B4020^B40_GE
OBX|12|NM|148066^MDC_ECG_V_P_C_RATE^MDC|1.5.1.2|0|264
864^MDC_DIM_BEAT_PER_MIN^MDC|||||R|||||||080019FFFE0B
4020^B40_GE
OBX|13|NM|131841^MDC_ECG_AMPL_ST_I^MDC|1.5.1.3|0.04|2
66419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE0B
4020^B40_GE
2062665-001 G