2 INSTALLATION

2.1 Accessories

The accessories supplied are as follows:
1 x reagent container
2 x calibration solution containers
3 x solution container sealing caps
1 x spares kit

2.2 Location

Install in a clean, dry, well ventilated and vibration-free location
giving easy access, and where short sample lines can be
used. Avoid rooms containing corrosive gases or vapors, e.g.
chlorination equipment or chlorine gas cylinders. It is also
advisable to have adjacent drains near ground level, so that
the waste outlet from the monitor can be as short as possible,
together with maximum fall. Power supplies should also be
adjacent. Ambient temperature within the range 5 C to 40 C.
2.3 Mounting – Fig. 2.1
See Fig. 2.1 for mounting procedure.
1
Mark the wall using the dimensions shown in
this figure.
Drill and plug the holes and screw in the top two
screws/bolts to leave a gap of 5mm between the
screw head and the wall.
Alternatively, with the enclosure carefully supported
against the wall, spot through using a suitable tool.
OUT
OF
SER
VICE
CAL
HOL
D
ALAR
M 1
ALAR
M 2
OUT
OF
SAM
PLE
3
Spot through lower fixing holes from backplate.
Remove enclosure before drilling lower fixing holes.
Note.
• Mains (power supply) and
signal cables are connected
through cable glands directly
into the electronic section.
• Sample and drain pipe work
are brought in through the
bottom of the case.
4

2.4 Sampling Requirements

In addition to being as close as possible to the monitor, the
sampling
representative sample. The sample must also conform to the
following conditions:
a) Sample flowrates must be greater than 5 ml min
b) Sample temperature should be within the range 5 C to 55 C.
c) Particles must be less than 10 mg l
740
2
Hook the
enclosure onto
the Enclosure
Hanger Bracket
screws.
55
85
Fig. 2.1 Mounting the Unit
point
must
provide
exceed 60 microns.
Note. To avoid erroneous readings and prevent
possible tube blockages the use of a sampling system
which provides the appropriate filtration of the sample is
recommended. ABB Ultrafiltration Units are the most
suitable for the purpose, details of which are available on
request. For power applications a 60 micron, in-line filter
should be used – part number: 0217 463.
Dimensions in mm.
Bracket Keyholes
6.5 o.d.
16.0 o.d.
110
347
Clearance
for lock
540
173.5
173.5
660
205.5
205.5
120 opening
411
6.3mm o.d.
Fixing Holes
a
thoroughly
mixed
-1
.
-1
and the size must not
M20 Cable
Gland Entry
43mm Centres
108.5
Lower

Mounting

Spacers
240
600
863 Ref

Advertisement

Table of Contents
loading

Table of Contents