6 INSTALLATION

6.1 Siting Requirements – Fig. 6.1

6.1.1 Transmitter

Notes.
• Mount in a location free from excessive vibration.
• Mount away from harmful vapours and/or dripping
fluids.
• Where possible, mount the transmitter at eye level to
allow an unrestricted view of the front panel displays
and controls.
Maximum Distance
750mm (29.5 in.)
Minimum Distance
200mm (7.88 in.)
A – Distance Between Analyzer and Sensor
0°C
(32°F)
Min.
B – Within Temperature Limits
NEMA 4X
C – Within Environmental Limits
Fig. 6.1 Siting Requirements
32
50°C
(122°F)
Max.
IP66

6.1.2 Sensor

Notes.
• The sensor is supplied fitted with mounting
brackets.
• Secure the sensor to a suitable vertical surface in a
location that enables easy access for maintenance
and calibration.