Test Outputs And Maintenance - ABB AV410 User Manual

Single and dual input dissolved organics monitor
Table of Contents

Advertisement

...5 PROGRAMMING

5.9 Test Outputs and Maintenance

-----
TEST/MAINTENANCE
-----
Test Outputs
4. 0 0
mA
20. 0
%
Test Output 1
-----
Maintenance
Auto.
On
-----
Off
Hold Outputs
28
Test Outputs
Displays the output test details for the analog outputs.
frame only is shown. The format of
Test Output 1
Continued below.
Maintenance
Test Output 1
The theoretical output current value.
Output current as a percentage of the full range current.
Use the
and
keys to adjust the displayed theoretical output current value to give
the output required.
Continued below.
Maintenance
See Section 7.1.
FACTORY SETTINGS
Advance to Test Output 2.
Test Output 2
Maintenance
Hold Outputs
Enables the relay action and analog outputs to be maintained.
Auto.
– Changes in relay action and analog outputs are disabled during sensor
calibration.
On
– Changes in relay action and analog outputs are disabled.
– Changes in relay action and analog outputs are not disabled.
Off
Note. The LEDs flash while the monitor is in 'Hold' mode.
Continued on next page.
Sensor A Outputs
See Section 7.1.
FACTORY SETTINGS
set to
Hold Outputs
Maintenance
set to
Hold Outputs
Automatic Time
Test Output 2
or
– return to main menu.
Off
On
– continued on next page.
Auto.
frame is identical.