Introduction; Principle Of Operation; Av400 Series Systems - ABB AV410 User Manual

Single and dual input dissolved organics monitor
Table of Contents

Advertisement

1 INTRODUCTION

1.1 Introduction
UV
Warning. The
contains a high intensity xenon strobe lamp that emits
ultraviolet (UV) radiation. This must
with the naked eye and must
while outside the sensor. Under normal operating
conditions, it is not possible to see the light source but,
if the sensor is dismantled with electrical power applied,
it may be possible to expose the eyes to the strobe
flash.
Many dissolved organic compounds (DOC) found commonly in
potable water absorb ultraviolet radiation. These include Humic
Acids, that gives water a characteristic yellow color, and
dissolved
organics,
that
Trihalomethanes (THMs).
The monitor is designed for use as a surrogate color monitor,
coagulation monitor/controller and to monitor for THM
precursors in potable water treatment plants.

1.2 Principle of Operation

The broad-spectrum, high intensity xenon strobe lamp, housed
in the emitter module, generates pulses of light that pass
through the sample water in the flowcell to a filtering and
detection system, contained in the receiver module. The
received light pulses are analyzed at two wavelengths; the
measurement wavelength of 254 nm and the reference
wavelength of 405 nm (at which the sample constituents of
interest do not absorb). This dual light path system provides
information that enables the measured value to be corrected for
any turbidity due to suspended matter in the sample. The
monitor is calibrated with a pure solution of a suitable organic
compound of known carbon content.
An automatic, microprocessor-controlled, dual-wiper system,
housed in the cleaner module, cleans the flowcell optical
windows periodically to ensure that the sensor remains
functional. Samples containing large solids and/or very high
concentrations of solids must be pre-filtered.
2
sensor
emitter
module
be viewed
NOT
be operated
NEVER
result
in
the
formation
1.3 AV400 Series Systems – Fig. 1.1
Note. An
AV400
configured as a matched system with each component
bearing the same serial number. If any part of a system
is replaced (transmitter or sensor[s]),
factory re-calibration must be carried out
Section 7.
The AV410 and AV411 low range monitors are primarily for
use in potable water applications, such as monitoring the
effectiveness of the coagulation control, THM precursor
detection and final treated water quality.
Range: 0 to 20mgl
of
The AV420 and AV422 high range coagulation monitors are
designed for use in potable water treatment plants to predict
the coagulant dose to be applied to the raw water and to
detect the rise in DOC from algal bloom toxins.
Range: 0 to 100mgl
7320 100
Low Range System
7320 200
High Range System
Fig. 1.1 AV400 Series Systems
System
is
supplied
a complete
-1
C maximum.
-1
C maximum.
AV400
AV400
factory-
– see

Advertisement

Table of Contents
loading

This manual is also suitable for:

Av411Av412Av420Av422

Table of Contents