...6 INSTALLATION
...6.5 Electrical Connections
6.5.2 Cable Entry Knockouts – Fig. 6.9
The monitor is supplied with 7 cable glands, one fitted and six to be fitted, as required, by the user – see Fig. 6.9.
Release the four captive
screws and remove
the terminal cover plate
Secure the cable
gland with the nut
40
1
6
Insert the cable gland into the hole
5
in the monitor case from the outside
Fig. 6.9 Cable Entry Knockouts
Factory-fitted cable gland
Place the blade of a small, flat bladed screwdriver
2
into the knockout groove and tap the
screwdriver smartly to remove the knockout
(see Caution below)
Caution. When
take great care not to damage wiring and
components within the transmitter.
Smooth the edges of the hole
3
with a small round or half-round file
4
Fit an 'O' ring seal to the the cable gland

Cable entry knockouts

removing
knockouts,