GE MEDICAL SYSTEMS
REV 5
ILLUSTRATION 10-4
CATHODE CABLE ROUTING
FROM TUBE
ILLUSTRATION 10-5
CATHODE CABLE
176.00
REF
Note:
2. From clamp K" the cable continues to strap point M" and then to clamp point
H." See Illustration 10-4.
CLAMP POINT H
STRAP POINT M
3. The cathode cable continues from clamp point H" and strap point G," to clamp
point E" to the front panel in a figure-eight configuration, and then terminates
at the transformer in the base of the unit.
D
The distances are referenced in Illustration 10-5.
CLAMP
CLAMP
POINT H"
POINT E"
STRAP
POINT G"
83.00
49.00
90.00
Strap point J" is installed with collimator cable.
10-5
(MODEL 2169360, 2236420 & 2275938 SERIES)
STRAP POINT G
CLAMP POINT E
STRAP
TUBE END
POINT M"
CLAMP
STRAP
POINT K"
POINT  J"
19.75
17.00
41.00
"0"
AMX-4+ SERVICE
DIRECTION 2173225-100