GE AMX-4+ Technical Manual page 28

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
Note:
ILLUSTRATION 1-8
FIELD LAMP ADJUSTMENT
9. Measure and record the horizontal A & B distances, in inches, between the x-
ray field and the light field. See Illustration 1-7.
10. Measure and record the vertical C & D distances, in inches, between the x-ray
field and the light field. See Illustration 1-6.
11. The total of the horizontal differences as well as the total of the vertical differ
ences must each be less than 1.8% of the SID:
D
A + B < 1.8% SID = A + B <0 .018  40 inches = A + B < 0.72 inches
D
C + D < 1.8% SID = C + D <0 .018  40 inches = C + D < 0.72 inches
If the above requirements are met, go to step 18. If they are not met, proceed with
the following steps.
Ideally, the x-ray field and the light field should coincide but this is rarely the
case. The best compromise is to move the light field so the x-ray field is cen
tered within it. Usually the light field area will be larger than the x-ray area.
12. To place the two fields as close to coincidence as possible, only the light field area
is moved. The light field area adjustment is made by repositioning the field lamp
housing.
13. Loosen the two Phillips head adjustment screws located in the back of the colli
mator. See Illustration 1-8.
14. Examine the last exposure and determine which direction to move the field lamp
to best position light field around the x-ray field.
15. Turn on the field lamp.
1-10
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
ADJUSTMENT SCREWS
AMX-4+ SERVICE