GE Healthcare
Direction 5743854-1EN, Revision 1
Section 4.0 Product Storage and Handling Requirements
4.1
Relative Humidity and Temperature
This section provides information for the environmental requirements for the storage of the system.
Table 1-2 Environmental Requirements (Relative Humidity & Temperature)
Product or Component
URP Detector
Wall Stand/Extended Wall Stand
Table (TBL)
OTS
System Cabinet(SKL1)
Maxiray 100-09 X-ray Tube(RAD)
Operator Console
PC Tower
LCD Monitor
Radiographic Stretcher Table (optional) 30%
Note: Table information is only applicable for Optima XR648.
Limits for rates of change:
In-Use
<10 degree C / hour
<30% / hour
STORAGE values only refer to equipment that is still in shipping containers. If the equipment is partially or
completely installed, refer to IN-USE values.
Relative Humidity
(Non-Condensing)
IN-USE
STORAGE
MIN
MAX MIN MAX
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
30%
60% 10% 85% 15°C (59° F) 35°C (95° F)
60% 10% 85% 15°C (59° F) 35°C (95° F)
Storage
<20 degree C / hour
<30% / hour
Chapter 1 - General Requirements
Optima XR642/XR648 Pre-Installation
Temperature
IN-USE
MIN
MAX
-5° C (23° F)
-20°C (-4° F)
-20°C (-4° F)
-20°C (-4° F)
-20°C (-4° F)
-20°C (-4° F)
-20°C (-4° F)
-20°C (-4° F)
-20°C (-4° F)
-20°C (-4° F)
STORAGE
MIN
MAX
50° C (122° F)
60° C (140° F)
60° C (140° F)
60° C (140° F)
60° C (140° F)
60° C (140° F)
60° C (140° F)
60° C (140° F)
60° C (140° F)
60° C (140° F)
Page 37