Introduction and technical data

1.1.

Purpose of manual

This document describes the detailed replacement procedure for the rotor control unit.
Additional stator inductance measurements are described to detect stator inductance problems which can lead
to rotor control failures.
1.2.

Items supplied

9890 000 0268x
Rotor Control HS
4512 984 28231
This manual
4512 116 02502
Service disk for replacement
1.3.

Compatibility

All OPTIMUS RAD/RF & C generators in combination with high-speed rotor control unit.
Rotor control HS
9890 000 0268x 4512 104 7360x / 7361x
Rotor control HS
9890 000 02212 4512 104 7140x
OPTIMUS RAD
9890 000 02001 release ≤ 2.x
OPTIMUS RAD/RF 9890 000 02001 release ≥ 3.x
OPTIMUS C
9890 000 02191 release ≥ 1.x
1.4.

Technical data

1.4.1. Tools
Tools
TC129
:
Tool kit, standard
Inductivity measuring device to be obtained locally
or order at SL Hamburg (4512 101 77141)
Service PC with AGenT version 2.1.2
1.5.

Safety information

The general legal and factory safety recommendations for this X-ray equipment and the following
recommendations must be strictly observed!
Start of installation, operation and maintenance work and especially electrical work must only be executed by
trained and authorized persons. This equipment must only be serviced by properly educated service specialists
who have received general and system-specific training as performed by Philips Medical Systems.
Warning!
The system/component must be switched OFF during replacement work.
Any X-ray unit produces ionizing radiation which may be harmful if not properly
controlled. Therefore, it is recommended that this equipment be operated in accordance
with the guidelines set down by the national council on radiation protection.
1.6.

Service reliability information

Permanently interested in quality improvement of the PMS products, we depend on getting information from the
field.
Therefore, please send us the current generator logfile information.
Please download the generator errorlog logfile.
The filename must contain the generator serial number and generator release.
e.g. "33960234.tdl" for rel. 3.3 and serial number 960234.
Send this file to:
Carsten Mais, Test & Integration Generators, PMS Hamburg,
as Attachment to E-Mail.