Periodical maintenance for this lab
washer is offered by Miele Professional
Technical Service.
After 1000 operating hours or at least
once a year is the recommended
interval.
The maintenance covers the following:
– Electrical safety
– Door mechanism and door seal
– Threads and connections in the
wash cabinet
– Water inlet and drain
– Internal and external dispensing
systems
– Spray arms
– Triple Filter System
– Drain pump and non-return faucet
– Steam condenser:
– All mobile units, baskets and inserts
If available:
– Connected printer
Part of the maintenance is a function
check:
– A program is run as a test
– A thermoelectrical measurement
– A leak tightness check
– All safety relevant measuring
systems (Display of faults)
Contact Miele Professional
Technical Service at 800-991-9380
for details.

Routine checks

The routine checks must be done daily
by the user before using the machine.
For the routine checks a checklist is
equipped with the washer.
Check the following points:
– the filters in the base of the wash
cabinet,
– the spray arms in the machine and at
the mobile units and inserts,
– the wash cabinet and the door seal,
– the mobile units, baskets and inserts,
and
– all safety equipment.