Connecting laser camera
•
Press the [CTRL] and [U] keys on the service PC simultaneously.
This opens the User Setup menu for the Memoskop.
•
Select the "C Documentation" sub-menu.
•
Configure monitor A for a SIREMOBIL Compact with one monitor.
•
Configure monitor B for a SIREMOBIL Compact with two monitors.
•
If the system is equipped with an area dose measurement device (Diamentor), open the
"K Documentation of Dose" menu and configure "yes".
Function tests
•
Connect an oscilloscope to the video input and the video output of the optical video
separator device. The video output must be terminated with 75 Ohm.
•
Open a test image (horizontal sawtooth) using the service PC.
The video signal should be 1:1.
Check for the following:
- correct termination (75 Ohm) of the video signals (cable)
- the sync amplitude and the total amplitude of the BAS signal
•
Connect a DMM to both contacts of the remote release button.
•
Switch the DMM to audible signal (continuity tester).
•
Press the hardcopy release button on the keyboard of the monitor trolley and the release
button on the basic system.
Wait approximately 7 seconds after pressing the first button before pressing the second.
Final steps
•
Switch off the SIREMOBIL.
•
Install all covers again.
•
Test the protective conductor using the protective conductor meter 44 15 899 RV 090.
Refer to TI 236 / ARTD - 002.731.17....
Changes to the previous version
Page 4, in the section, paragraph "When a video printer is present"
text "green and white ..." inserted
TD SP 2 / Arnold
Siemens AG
Medical Engineering
The DMM must beep briefly.
Register 3
TD SP 1 / Schlee
RXR2-130.033.02
Rev. 02
03.97
SMS Iselin / Weiss Costa
Page 7 of 8
SIREMOBIL Compact
TD SP 1
7
1
1
1